OPTIONAL FUNCTION; NOT INSTALLED
Error Number
The function you requested requires a capability provided by an option to the
1
standard analyzer. That option is not currently installed. (Refer to
Options Available"
the available options.)
OVERLAP!LIST TYPE CHANGED TO STEPPED
Error Number
The list type changed to stepped because one or more frequency segments in the
211
swept list table overlapped. Change the frequency ranges of the overlapping
segments and switch back to swept list mode.
OVERLOAD ON INPUT A, POWER REDUCED (ES only)
Error Number
See OVERLOAD ON INPUT R, POWER REDUCED (error number 57).
58
OVERLOAD ON INPUT B, POWER REDUCED (ES only)
Error Number
See OVERLOAD ON INPUT R, POWER REDUCED (error number 57).
59
OVERLOAD ON REFL PORT, POWER REDUCED (ET only)
Error Number
See OVERLOAD ON INPUT R, POWER REDUCED (error number 57).
58
OVERLOAD ON TRANS PORT, POWER REDUCED (ET only)
Error Number
See OVERLOAD ON INPUT R, POWER REDUCED (error number 57).
59
OVERLOAD ON INPUT R, POWER REDUCED
Error Number
You have exceeded approximately +14 dBm at one of the test ports. The RF
output power is automatically reduced to −85 dBm. The annotation P⇓ appears
57
in the left margin of the display to indicate that the power trip function has been
activated. When this occurs, reset the power to a lower level, then toggle the
softkey to switch on the power again.
PARALLEL PORT NOT AVAILABLE FOR GPIO
Error Number
You have defined the parallel port as COPY for sequencing in the HP-IB menu.
165
To access the parallel port for general purpose I/O (GPIO), set the selection to
PARALLEL [GPIO]
in
Chapter 7 , "Options and Accessories"
.
Error Messages
Error Messages in Alphabetical Order
"Analyzer
for a description of
6-15