HP 8719ET Reference Manual page 19

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

Table 1-7 3.5-mm Device Connector Type
Network Analyzer: HP 8719ES/20ES, Option 007
Calibration Kit: HP 85052D (3.5-mm 50 Ω)
Cables: HP 85132F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
0.05 to 0.5 GHz
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
±(0.006 + .02/°C)
Magnitude (dB)
±(0.038 + 0.1/°C)
Phase (deg)
Trans. Tracking
±(0.018 + .02/°C)
Magnitude (dB)
±(0.118 + 0.1/°C)
Phase (deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8719/20ES)
0.5 to 2 GHz
42
42
37
37
42
41
±(0.010 + .03/°C)
±(0.069 + 0.1/°C)
±(0.019 + .03/°C)
±(0.123 + 0.1/°C)
Specification
2 t0 8GHz
38
30
36
±(0.030 + .03/°C)
±(0.200 + 0.3/°C)
±(0.080 + .03/°C)
±(0.531 + 0.3/°C)
8 to 20 GHz
36
28
34
±(0.031 + .04/°C)
±(0.205 + 0.5/°C)
±(0.141 + .04/°C)
±(0.928 + 0.5/°C)