HP 8719ET Reference Manual page 22

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8719/20ES)
Table 1-10 Type-N Device Connector Type
Network Analyzer: HP 8719ES/20ES, Standard
Calibration Kit: HP 85054D (Type-N, 50 Ω)
Cables: HP 85132F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
±(0.006 + .02/°C)
Magnitude (dB)
±(0.040 + 0.1°C)
Phase (deg)
Trans. Tracking
±(0.031 + .02/°C)
Magnitude (dB)
±(0.205 + 0.1/°C)
Phase
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
1-12
50 to 500 MHz
0.5 to 2 GHz
40
38
40
±(0.006 + .03/°C)
±(0.040 + 0.1/°C)
±(0.033 + .03/°C)
±(0.218 + 0.1/°C)
Specification
2 to 8 GHz
40
36
38
33
40
36
±(0.009 + .03/°C)
±(0.059 + .30/°C)
±(0.094 + .03/°C)
±(0.620 + 0.3/°C)
8 to 18 GHz
34
29
34
±(0.027 + .04/°C)
±(0.178 + 0.5/°C)
±(0.168 + .04/°C)
±(1.109 + 0.5/°C)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8720etHp 8719esHp 8722etHp 8720es8722es

Table of Contents