HP 8719ET Reference Manual page 29

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

Table 1-16 3.5-mm Device Connector Type
Network Analyzer: HP 8722ES, Option 400
Calibration Kit: HP 85052C (3.5-mm, 50 Ω)
Cables: HP 85131F
Calibration: TRL
IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
±(0.006 + .03/°C)
Magnitude (dB)
±(0.040 + 0.1/°C)
Phase (deg)
Trans. Tracking
±(0.017 + .03/°C)
Magnitude (dB)
±(0.112 + 0.1/°C)
Phase(deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
0.05 to 2 GHz
2 to 8 GHz
48
40
48
±(0.005 + .03/°C)
±(0.033 + 0.3/°C)
±(0.013 + .03/°C)
±(0.086 + 0.3/°C)
HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8722ES)
Specification
8 to 20 GHz
50
50
50
50
50
50
±(0.005 + .04/°C)
±(0.033 + 0.5/°C)
±(0.016 + .04/°C)
±(0.106 + 0.5/°C)
20 to 26.5 GHz
50
50
50
±(0.005 + .06/°C)
±(0.033 + 1.0/°C)
±(0.023 + .06/°C)
±(0.152 + 1.0/°C)
1-19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8720etHp 8719esHp 8722etHp 8720es8722es

Table of Contents