Advertisement

MEMORY FOR CURRENT SEQUENCE IS FULL
Error Number
All the memory in the sequence you are modifying is filled with instrument
132
commands.
MORE SLIDES NEEDED
Error Number
When you use a sliding load (in a user-defined calibration kit), you must set at
71
least three slide positions to complete the calibration.
NO CALIBRATION CURRENTLY IN PROGRESS
Error Number
The
69
already in progress. Start a new calibration.
NO DISK MEDIUM IN DRIVE
Error Number
You have no disk in the current disk unit. Insert a disk, or check the disk unit
41
number stored in the analyzer.
NO FAIL FOUND
Service Error
The self-diagnose function of the instrument operates on an internal test failure.
Number 114
At this time, no failure has been detected.
NO FILE(S) FOUND ON DISK
Error Number
No files of the type created by an analyzer store operation were found on the
45
disk or the disk drive is empty. If you requested a specific file title, that file was
not found on the disk.
NO IF FOUND: CHECK R INPUT LEVEL
Error Number
The first IF signal was not detected during pretune. Check the front panel R
5
channel jumper. If there is no visible problem with the jumper, refer to the
service guide for troubleshooting.
NO LIMIT LINES DISPLAYED
Error Number
You can turn limit lines on but they cannot be displayed on polar or Smith chart
144
display formats.
NO MARKER DELTA -SPAN NOT SET
Error Number
You must turn the delta marker mode on, with at least two markers displayed,
15
in order to use the
RESUME CAL SEQUENCE
MARKER → SPAN
Error Messages in Alphabetical Order
softkey is not valid unless a calibration is
softkey function.
Error Messages
6-13