Table 1-22 2.4-mm Device Connector Type
HP 8722ES (2.4-mm, 50 Ω)
Description
a
Directivity
a
Source Match
Standard/
Option 400 (dB)
Option 007 or
Option 085 (dB)
a
Load Match
Standard/
Option 400 (dB)
Option 007 or
Option 085 (dB)
Reflection
b
Tracking
(dB)
Transmission
b
Tracking
Tracking Stability
(Ratio
Measurement)
Magnitude (dB)
Phase (dB)
c
Crosstalk
a. Does not include the effect of the cable set on the test ports.
b. Excludes rolloff below 500 MHz, which is typically −18 dB at 100 MHz, and −25 dB at
50 MHz.
c. Measurement conditions: Normalized to a thru, measured with two shorts, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of the
maximum power out or the maximum receiver power.
d. Limited by noise floor.
50 to
0.5 to
500 MHz
2 GHz
23
23
17
17
17
17
18
18
21
21
±3
±3
±3
±3
0.02/°C, typ.
0.03/°C, typ.
0.1/°C, typ.
0.1/°C, typ.
d
85
60
HP 8719/20/22ES Specifications and Characteristics
Characteristic
2 to
8 GHz
20 GHz
21
12
15
15
17
±3
±3
0.03/°C, typ.
.04 dB/˚C, typ.
0.3/°C, typ.
0.5 ˚/˚C, typ.
85
Instrument Specifications
8 to
20 to
40 GHz
16
15
11
7
11
8
12
10
13
10
±3
±6
±3
±6
.06 dB/˚C, typ.
1.0 ˚/˚C, typ.
82
72