HP 8719ET Reference Manual page 165

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

Hardkey/Softkey Reference
Analyzer Functions
ONE-PATH 2-PORT
ONE SWEEP
OP PARMS (MKRS etc)
OPEN
OPEN (F)
OPEN (M)
P MTR/HPIB TO TITLE
PARALL IN BIT NUMBER
PARALL IN IF BIT H
5-38
leads to the series of menus used to perform a
high-accuracy two-port calibration without an
S-parameter test set. This calibration procedure
effectively removes directivity, source match, load match,
isolation, reflection tracking, and transmission tracking
errors in one direction only. Isolation correction can be
omitted for measurements of devices with limited dynamic
range. (The device under test must be manually reversed
between sweeps to accomplish measurement of both input
and output responses.) The required standards are a
short, an open, a thru, and an impedance-matched load.
This mode does not measure each sweep, but corrects each
point with the data currently in the power meter
correction table.
provides a tabular listing on the analyzer display of the
key parameters for both channels. The screen menu is
presented to allow hard copy listings and access new pages
of the table. Four pages of information are supplied. These
pages list operating parameters, marker parameters, and
system parameters that relate to control of peripheral
devices rather than selection of measurement parameters.
defines the standard type as an open, used for calibrating
reflection measurements. Opens are assigned a terminal
impedance of infinite Ωs, but delay and loss offsets may
still be added.
for cal kits with different models for male and female test
port standards, this selects the open model for a female
test port. Note that the sex of a calibration standard
always refers to the test port.
for cal kits with different models for male and female test
port standards, this selects the open model for a male test
port. Note that the sex of a calibration standard always
refers to the test port.
gets data from an HP-IB device set to the address at which
the analyzer expects to find a power meter. The data is
stored in a title string. The analyzer must be in system
controller or pass control mode.
while creating a sequence, this softkey will insert a
command that selects the single bit (0 to 4) that a
sequence will be looking for from the GPIO bus.
while creating a sequence, this softkey inserts a command
to jump to another sequence if the single input selected is
in a high state.