HP 8719/20/22ES Specifications and Characteristics
Instrument Specifications
Table 1-35 Test Port Input
HP 8722ES Test Port Input
Dynamic Accuracy (Characteristic)
For input ports 1 and 2, accuracy of the test port input power reading relative to the reference
input power level.
• Inputs: testport 1 and 2
• For test port powers > −50 dBm and < 0 dBm, magnitude dynamic accuracy is
0.020 dB + 0.0015 dB/dB from the reference power, phase dynamic accuracy is
0.132 deg + 0.0066 deg/dB from the reference power. For test port powers > −80 dBm and
> −50 dBm, magnitude dynamic range is .02 dB + .003 dB/dB from the reference power.
• For test port powers up to 0 dBm.
1-38