REFLECT AND LINE
REFLECTION
REFLECTION 1-PORT
REMOVE ADAPTER
RENAME FILE
RESET COLOR
RESPONSE
RESPONSE ISOL'N
measures the reflection and thru paths of the current
calibration standard.
leads to the reflection calibration menu.
leads to the reflection calibration menu. (ET only)
completes the adapter removal procedure, removing the
effects of the adapter being used.
allows you to change the name of a file that has already
been saved.
resets the color being modified to the default color.
• When in the specify class more menu,
used to enter the standard numbers for a response
calibration. This calibration corrects for frequency
response in either reflection or transmission
measurements, depending on the parameter being
measured when a calibration is performed. (For default
kits, the standard is either the open or short for
reflection measurements, or the thru for transmission
measurements.)
• When in the response cal menu,
the frequency response calibration. This is the simplest
and fastest accuracy enhancement procedure, but
should be used when extreme accuracy is not required.
It effectively removes the frequency response errors of
the test setup for reflection or transmission
measurements.
• When in the specify class more menu,
RESPONSE ISOL'N
numbers for a response and isolation calibration. This
calibration corrects for frequency response and
directivity in reflection measurements, or frequency
response and isolation in transmission measurements.
• When in the response and isolation menu,
RESPONSE ISOL'N
perform a response and isolation measurement
calibration, for measurement of devices with wide
dynamic range. This procedure effectively removes the
same frequency response errors as the response
calibration. In addition, it effectively removes the
isolation (crosstalk) error in transmission
measurements or the directivity error in reflection
measurements. As well as the devices required for a
simple response calibration, an isolation standard is
required. The standard normally used to correct for
Hardkey/Softkey Reference
Analyzer Functions
RESPONSE
RESPONSE
is used to enter the standard
leads to the menus used to
is
leads to
5-47