7 Maintenance and disinfection

WARNING
The device should be cleaned and disinfected before it is used for the first time, and also
between patients.
The disinfection of the ventilator must be carried out as stated in the manufacturer's instructions
and in accordance with the current protocol of the health care institution.
Service personnel must be familiar with cleaning and disinfection protocols as well as the
precautions required for certain materials.
The following figure shows the components which need to be cleaned and/or replaced:
Oxygen
sensor
Dust filter

7.1 Maintenance schedule

Any recommendations concerning cleaning frequency and the replacement of disposable
components are provided by the manufacturers of those components. As an indication, this table
provides information on the components that must be cleaned before reuse.
Patient circuit
Antibacterial filter
Expiratory valve for double circuit
Dust filter
Nebuliser
Oxygen sensor
Device casing
NOT013281-8
Ventilator casing
Patient circuit
Expiratory valve for double
circuit
For the same patient
Hospital protocol or manufacturer
If used for antiviral and antibacterial
protection, refer to the manufacturer's
instructions
Hospital protocol or one week
Hospital protocol or manufacturer
Refer to ventilator alarm if necessary
Hospital protocol or one week
Antibacterial filter
Clapper
+
Maintenance frequency
Between patients
Yes
Yes
With antibacterial filter: hospital
protocol or one week
Without filter: clean between patients
Every 6 months at least
Yes
Yes
Maintenance and disinfection
81/102