Troubleshooting - ResMed Elisee 350 Clinical Manual

Table of Contents

Advertisement

6.6 Troubleshooting

WARNING
If, after trying these solutions, the activated alarm persists, turn the device off and contact
your technician.
Alarms that cannot be stopped
Symbol
Alarms that can be stopped temporarily (muted for 120 seconds)
Symbol
L
L
78/102
Message
Priority
INT BATT EMPTY
High
MINI FIO
High
2
MAXI FIO
High
2
NiMH BATTERY
Low
Message
Priority
O
FAULT
High
2
PATIENT APNEA
Medium
EXT BATT EMPTY
Medium
EXT BATT LOW
Low
INT BATT LOW
Medium
CIRCUIT OPEN
High
TEST FAILED
High
MAXI F
Low
MAXI LEAKS
High
WRONG CIRCUIT
High
LOW PEEP
Medium
MAXI P EXPI
High
Cause
Internal battery charge less than
5%
Supplied FiO
too low
2
Supplied FiO
too high
2
An unknown external battery
pack has been connected
Cause
HP O
supply missing
2
No spontaneous cycles were
inititated during Tapnea starting
from the end of the previous
expiratory phase
External battery charge less
than 5%
External battery charge less
than 15%
Internal battery charge less than
15%
Patient circuit is poorly
connected
Patient circuit is faulty
Manual test failed
Respiratory rates exceed the
upper threshold of the
Respiratory Rate alarm
Leaks exceed the upper
threshold of the High Leak
alarm
Type of patient circuit used is
different from the one
determined by the test
The PEEP delivered to patient is
lower than the target PEEP
The expiratory pressure is too
high compared with the target
PEEP
Solution
Change the power supply
Check that an oxygen supply is connected and
O
2
press the
100%
button (see page 65)
2
min
Check that an oxygen supply is connected and
O
2
press the
100%
button (see page 65)
2
min
Connect a compatible external battery pack
(Li-Ion type)
Solution
Connect the high-pressure O
tube
2
Check for leaks from the mask or tubing.
If necessary, adjust the apnea time.
For the double circuit only, adjust the trigger
value
Connect the device to mains supply or change
the external battery pack
Connect the device to mains supply or change
the external battery pack
Change the power supply
Reconnect the patient circuit
Change the patient circuit
Repeat the manual test
Check that Respiratory Rate alarm and
inspiratory trigger settings are correct
Check that the High Leak alarm threshold
setting is correct and check the integrity of
the patient circuit
Check the circuit and repeat the test
Check the integrity of the patient circuit
Check the expiratory valve control and check
the patient circuit is not blocked
NOT013281-8