5.4 Recommendations for use during transport

During transport, we recommend using the ventilator in its ergonomic transport bag, which must
be firmly attached to the vehicle.
Before using the ventilator during transport, connect it to mains power for long enough to ensure
that its internal and/or external batteries are charged.
In the hospital vehicle, before connecting the ventilator to a patient:
1. Check that the ventilator is in its transport bag, and that the bag is installed in the transport
bracket system or any other secure attachment system according to the current legislation in
your country.
2. Ensure that the device is powered (using one or more batteries, or an external power supply)
in order to minimise temperature variations.
3. Perform a manual test.
4. Check that the compartment under the device is not empty; if necessary, insert the dummy
pack.
5. Ensure that an additional O
Note: In accordance with the standard EN 794-3, it is recommended that the user has a means
of backup ventilation available.
CAUTION
During transport, the mains power pack must not be in the compartment under the
device, but may be connected to the ventilator through the external DC power socket.
Do not use the ventilator in dangerous, explosive, or biologically contaminated
situations.

5.5 Configuring ventilation

The ventilator settings can be accessed from the main screen. There are three options:
TEST
TEST
50/102
Setting a new configuration (all parameter settings are restored to their default values)
Selecting the most recently used configuration
Starting ventilation using one of five preset programs (see "Pre-programming settings" on
page 54).
Last test the 18/06/10
Double cirtcuit : PASS
NEW
NEW
PRESET
PRESET
SETTING
SETTING
SETTING
SETTING
PREVIOUS SETTING
PREVIOUS SETTING
ACV
ACV
bottle is available.
2
Press to start ventilation using a preset program
Press to set a new configuration
Press to select the most recently used configuration. The icons indicate its main
characteristics, for example:
: Invasive ventilation
: Adult
ACV mode