Alarm Settings - ResMed Elisee 350 Clinical Manual

Table of Contents

Advertisement

Relative alarm priorities
Highest priority
+
-
Lowest priority

6.4 Alarm settings

CAUTION
If you change any of the alarm settings, see Section 6.5 on page 74 for instructions on
testing the alarms.
: Non-invasive ventilation;
Value:
Patient type:
Max. resp. rate
(bpm)
Max. pressure
(cm H
O)
2
°
Maxi
e (L/min)
°
Mini
e (L/min)
°
Maxi
i (L/min)
°
Mini
i (L/min)
Maxi Vte (mL)
NOT013281-8
High-priority alarms
LP – Low Flow?
LP – CIRCUIT OPEN
MAXI P inspi
O
FAULT
2
MAXI LEAKS
MAXI P expi
WRONG CIRCUIT
MINI V/MIN
MINI Vt
MAXI Vt
FIO
21
2
MAXI FIO
2
MINI FIO
2
PRESENCE O
HP
2
INT BATT EMPTY
TEST FAILED
: Invasive ventilation
Default value
Child
Adult
Child
35
15
30
40
5
15
1.0
3
5.0
15
1.0
3
50
750
1000
200
550
Medium-priority alarms
EXT BATT EMPTY
INT BATT LOW
LOW PEEP
PATIENT APNEA
Settings range
Minimum
Maximum
Adult
Child
10
5
3.5
19.5 / NO
0.2
10.0
3.5
19.5 / NO
0.2
10.0
200
600
750 / NO
1000
Low-priority alarms
NiMH battery
EXT BATT MISSING
EXT BATT LOW
EXT MISSING
MAINS MISSING
EXPI resist
INSPI resist
MAXI V/MIN
MAXI F
ADJUST CLOCK
VENTIL STOPPED
REPLACE BLOWER
REPLACE O
CELL
2
Increment
Adult
99
1
99
1
39.5 / NO
0.5
20.0
0.2
39.5 / NO
0.5
20.0
0.2
10
1000
50
2400 / NO
100
Alarms and troubleshooting
73/102