6.2 Alarm types

Type 1: Ventilation monitoring alarms
If two alarms with different priority levels are triggered at the same time, the higher priority alarm
takes precedence over the lower priority alarm. If two alarms with the same priority level are
triggered at the same time, refer to the table under the heading "Relative alarm priorities" on
page 73.
Note: The 'Trigger threshold [can be stopped?]' column in the following table indicates for each
alarm whether it can be stopped permanently (YES), stopped temporarily (2 min) or not stopped
at all (NO).
Symbol
L
L
L
L
L
L
70/102
Message
Alarm
MINI Vt
Minimum Expired Tidal
Volume
(double circuit)
MAXI Vt
Maximum Expired Tidal
Volume
(double circuit)
MINI Vt
Minimum Inspired Tidal
Volume
(single circuit)
MAXI Vt
Maximum Inspired Tidal
Volume
(single circuit)
MINI V/min
Minimum Expired Minute
Volume
(double circuit)
MAXI V/min
Maximum Expired Minute
Volume
(double circuit)
MINI V/min
Minimum Inspired Minute
Volume
(single circuit)
MAXI V/min
Maximum Inspired Minute
Volume
(single circuit)
MAXI P
Maximum Inspiratory
INSPI*
Pressure
MAXI P EXPI*
Maximum Expiratory Pressure
LP – Low
Minimum Pressure (volume-
flow?
controlled modes)
Double circuit
CIRCUIT
Minimum Pressure (volume-
OPEN
controlled modes)
Double circuit
CIRCUIT
Minimum Pressure (volume-
OPEN
controlled modes)
Single circuit
CIRCUIT
Minimum Pressure (pressure-
OPEN
controlled modes)
CIRCUIT
Minimum pressure (pressure-
OPEN
controlled modes, CPAP)
LOW PEEP
Minimum Expiratory Pressure
(double circuit)
MAXI F
Maximum Respiratory Rate
Trigger
Priority
threshold [can
be stopped?]
High
Alarm setting
[2 min]
High
Alarm setting
[2 min]
High
Alarm setting
[2 min]
High
Alarm setting
[2 min]
High
Alarm setting
[2 min]
Low
Alarm setting
[2 min]
High
Alarm setting
[2 min]
Low
Alarm setting
[2 min]
High
Alarm setting
[2 min]
High
10 mbar + PEEP
[2 min]
High
3 mbar + leak
< 80%
[2 min]
High
3 mbar + leak
≥ 80% [2 min]
Δ
High
Pi alarm
threshold
[2 min]
High
(Pi or PS) – 2 mbar
at the end of
inspiration
[2 min]
High
(CPAP+1)/3 during
inspiration [2 min]
Medium
Alarm setting
[2 min]
Low
Alarm setting
[2 min]
Time to trigger
5 consecutive cycles above
alarm threshold
5 consecutive cycles above
alarm threshold
5 consecutive cycles above
alarm threshold
5 consecutive cycles above
alarm threshold
Immediately
Immediately
60 s
60 s
3 consecutive cycles above
alarm threshold
300 ms above alarm threshold
2 complete cycles below the
threshold or maximum of 17 s
2 complete cycles below the
threshold or maximum of 17 s
2 complete cycles below the
threshold or maximum of 17 s
2 complete cycles below the
threshold or maximum of 17 s
2 complete cycles below the
threshold or maximum of 17 s
30 cycles below alarm
threshold
4 consecutive cycles above
alarm threshold
NOT013281-8