2.4 Power supplies

The Elisée 350 can be used with several different power supplies. The device automatically
selects the power supply to be used, according to the following hierarchy:
1. A mains power pack installed in the compartment under the device.
2. An external power supply connected to the Elisée external DC power socket, which could be:
A mains power pack located ouside the ventilator
An external power supply (12
3. An external battery pack installed in the compartment under the device.
4. The internal battery.
External battery
pack (blue label,
no connectors)
CAUTION
Only power supplies or power cords designed and distributed by ResMed should be
connected to the ventilator.
Note: In the event of a loss of mains power, an audible and visual alarm is activated. It can be
stopped by pressing the Alarm Silence button.
NOT013281-8
Product code (PACXXX)
28 V DC) via the cigarette lighter.
Internal battery
4
3
1
Mains power pack (grey label and
two connectors)
Removable dust filter
Speaker
Ventilator serial number label. E.g. EL2H1101001:
Product
Year of
code
manufacture
EL2H
11
Compartment for mains power pack or external battery
pack
External power supplies
Mains power pack with external power cord (in this case, the
pack is treated as an external power supply)
Month of
Manufacture
manufacture
No.
01
001
2
External power
supply (via vehicle
cigarette lighter)
Introduction
9/102