Calibrating the oxygen sensor to 21% FiO
This function allows you to manually calibrate the oxygen sensor to 21% by passing ambient air
through the sensor.
cm
0 0
H
O
2
90
80
70
60
50
40
┘HP
30
20
10
0
<PE
cm
0
H
O
2
90
80
70
60
50
40
HP
30
20
10
<PE
0
2 Low-pressure oxygen
Before using low-pressure oxygen, check that:

5.11 Using the nebuliser

Before using the nebuliser, check that:
66/102
ACV
ACV
O
2
100%
Freeze->Touch curves
2
min
info
P
16
0
v
28
0
V
V
FiO
FiO
P
peak
peak
Ti Ti
Te
Te
2
cm
21
%
20,3
501
503
1/4
ml
ml
H
O
2
O
Consumed:
Consumed:
- - l/min
- - l/min
2 2
MENU
MENU
Configuration
Options ...
Back
Locked touch screen
New configuration
Stop machine
ACV
ACV
Ventil.
Ventil.
stop
stop
Select a menu
Back
info
Nebul.
Nebul.
Config.
Config.
Calibr.
Calibr.
Setting
Setting
O
2
The oxygen sensor is installed in the ventilator (see "Installing the oxygen sensor" on
page 38)
The manual test was successful (which means that the oxygen sensor is working).
See "3 Manual test" on page 48
The "Low pressure oxygen input" parameter is set to YES (see "Low-pressure oxygen
input" on page 63).
The oxygen sensor is installed in the ventilator (see "Installing the oxygen sensor" on
page 38)
The manual test was successful (which means that the oxygen sensor is working).
See "3 Manual test" on page 48
The high-pressure oxygen equipment is connected to the ventilator (see "High-pressure
oxygen" on page 39)
The "Low pressure oxygen input" parameter is set to NO (see "Low-pressure oxygen
input" on page 63).
O
2
M o
1. Press
100%
to supply 100% FiO
2
min
d e
s
corresponding icon will flash.
P a
r a
m
O
2
2. Press
100%
again to interrupt the supply before the end of
2
min
the two minutes.
Note: You cannot interrupt the oxygen supply during the first
four ventilation cycles, because this is the time required to
calibrate the sensor.
1. After selecting one of the configuration types (New, Most
Recent or Preset) on the main screen, press the
the right-hand side of the device.
2. Press
.
MENU
The following screen is displayed.
3. Press "Calibr. O
" to start the calibration.
2
for two minutes. The
2
2
button on
NOT013281-8