3.3 Ventilation settings and measurements

Ventilation parameters must be determined and set under the supervision of a physician or other
competent and trained personnel.
Ventilation settings
Adult ventilation settings
Parameter
FiO
(%)
2
F (bpm)
Ti (s)**
I:E**
Ti Max (s)
Tapnea (s)
Pi (cm H
O)
2
PS (cm H
O)
2
PEEP (cm H
O)
2
CPAP (cm H
O)
2
V
(ml)
T
Max flow (L/min)
°
max**
Rise time
Tplateau (s)
Flow shape
Recruitment period (min)
Recruitment size (s)
NOT013281-8
E
I
E
I
E
Default
INVASIVE
value
Min.
21
60
30
15
2
1
0.3
1:2.0
1:0.4
1.2
1
15
10
15
5
8
5
0
0
8
3
500
300
1000
60
10
1
1
0
0
1
1
NO
1/NO
2
0.3
3
Set CPAP
t (s)
t (s)
Ventilation type
NON-INVASIVE
Max.
Min.
Max.
30
21
30
100
30
100
50
2
50
3
0.3
3
1:9.9
1:0.4
1:9.9
3
1
3
60
10
60
60
5
60
60
5
60
25
0
25
25
3
25
1000
300
1000
2500
1000
2500
120
10
120
4
1
4
2
0
2
5
1
5
60
1/NO
60
3
0.3
3
40
3
40
Setting
Precision of
Increment*
set values
9
± 3
5
1
±
1
0.1
± 0.1
1:0.1
0.1
1
±
1
1
±
5%
1
±
5%
1
±
0.5 cm H
O
2
for values of
1–5 cm H
O,
2
10% for higher
values
1
±
10
10% or 10 ml
100
1
1
0.1
± 0.1
1
1
0.1
± 0.1
1
Ventilation
29/102