Run Method Screen 2 - Agilent Technologies 850-DS Operator's Manual

Dissolution sampling station
Table of Contents

Advertisement

4
Operating the 850-DS

Run Method Screen 2

Figure 22.
This screen allows you to enter parameters that are stored with the method run.
These parameters are available on a printout if your unit has the optional printer
installed.
Option
Operator
Lot
Batch
Note
Dissolution
Apparatus Control
Instant
Vessel Temperature
Bath Temperature
50
Method Start Parameters
Function
25-character field for operator name entry.
25-character field for lot information.
25-character field for batch information.
25-character field for additional notes.
Allows the Agilent 850-DS to control a dissolution apparatus.
Note: Enable this option only if the Agilent 850-DS is connected to another
Agilent/Varian/VanKel dissolution apparatus. If connected to a third-party
dissolution apparatus with resident probes, do not check this box.
Start a method instantly.
Will start a method once the vessels have achieved a stable temperature as
specified in the method to be run within the temperature tolerance settings of
the 850-DS.
Will start a method once the bath has achieved a stable temperature as
specified in the method to be run within the temperature tolerance settings of
the 850-DS.