Initial Power Up; Sample Tray Installation - Agilent Technologies 850-DS Operator's Manual

Dissolution sampling station
Table of Contents

Advertisement

3
Setting Up the 850-DS

Initial Power Up

1 Make sure there are no obstructions inside the sample tray area of the Agilent
2 Make sure the door is fully closed.
3 Turn on power using the front panel power switch. The unit will execute a self
4 Verify the unit passed the initialization and is at the Main screen.
NOTE

Sample Tray Installation

1 From the 850-DS main screen, press Eject to eject the sample tray platform.
C A U T I O N
2 Load the 12 x 8 sample tray (for 16 x 100 mm test tubes) onto the platform.
3 Press Eject. The sample tray platform retracts the sample tray to the home
24
850-DS
test and initialization.
Before operating the Agilent 850-DS, a sample collection tray with
suitable test tubes, vials or well plates must be inserted. Failure to
do so may cause liquids to be pumped into the base of the unit.
The sample tray platform moves to the front of the 850-DS. The movement
forward will pause and the unit will beep to alert the user that the tray platform
will extend outside the unit.
Do not attempt to insert or remove the sample tray until
movement of the platform has completed.
position in the 850-DS.
850-DS Operator's Manual