4
Operating the 850-DS

Sample Collection Parameters (Apparatus 1/2/5/6)

Figure 13.
Option
Sample Volume
Prime Volume
Purge Volume
Waste Drop Volume
Dual Sample
Estimated Min
Transfer Time
Prefill Tubes / Vials
Media Change
40
Sample Collection Parameters - Screen 1
Function
Volume to be collected into the sample test tubes / vials.
Volume used to prime all of the sample collection lines prior to collecting a
sample.
Volume used to purge all of the sample collection lines after collecting a sample.
Volume used to rinse the sample needles prior to delivering each sample. This
volume, in addition to the sample volume, should be accounted for if media
replacement is enabled.
When enabled, collects two consecutive samples for each timepoint. Uses the
sample volume specified above. This option also halves the total number of
timepoints available.
This is a calculated estimate, based on the previously entered sample collection
parameters, of how frequently samples may be taken.
Use this option to fill each test tube or sample vial with the specified volume (in
mL). This volume will be taken from the dissolution vessels and automatically
replaced using the Media Replacement channel. The number of rows filled is
based on the number of time points of the method. Ensure the media
replacement line is connected and sufficient volume is available in the container
to maintain initial vessel volume.
When enabled, allows you to manually intervene for a media modification.
850-DS Operator's Manual