Canon imageFORMULA DR-X10C Service Manual page 107

Hide thumbs Also See for imageFORMULA DR-X10C:
Table of Contents

Advertisement

4) Place the reading holder
move the screw
black) and remove the side plate
same way, remove the side plate on the
opposite side. Slide the reading holder,
and while unhooking the fitting parts
pairs), remove it.
Note: The right side plate differs from the left
side plate in shape. The right side plate
shown below has a stopper
Figure 3-517
5) Lift the CIS unit
both flat cables
the unit.
Figure 3-518
COPYRIGHT © CANON ELECTRONICS INC. 2008
1
upward, re-
2
(M3, self-tapping,
3
. In the
4
5
.
1
, and while removing
2
from the slit, remove
CANON DR-X10C FIRST EDITION
CHAPTER 3 DISASSEMBLY & REASSEMBLY
6) Remove 2 flat cables
nectors. While unhooking the fitting parts,
remove 2 grounding plates
Note: The connector has a lock. Release the
lock according to the previous section,
(7
"2. Reading PCB".
Notes on assembling
The grounding plates are same, but as-
semble them by aligning vertical and
horizontal directions with original positions.
Do not forget to mount a guide ring on the
shaft on the back side of the CIS unit.
1
from the con-
2
.
Figure 3-519
3-37