Chapter 2 Service Mode Access - Excell PW Service Manual

Ip67 waterproof
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD

CHAPTER 2 SERVICE MODE ACCESS

Set the jumper SWA1 on the main board to the ADJ position (EEPROM UNLOCKED).
n
Switch the PW on. The display shows
When finished, set the jumper SWA1 back to the LOCK position.
n
If the jumper SWA1 is returned to the LOCK position during calibration, the PW exits the
n
service mode automatically.
Press
ON/ZERO
⇒ Capacity Configuration
⇒ Weighing Calibration
⇒ Linearity Calibration
PW Service Manual
Press
key
ON/ZERO
Press
key
ON/ZERO
Press
key
ON/ZERO
Press
key
ON/ZERO
key
.
Press
Press
UNITS
Press
UNITS
Press
UNITS
Press
UNITS
10
key
UNITS
key
key
key
key
⇒ Function Setting
⇒ Gravity Setting
SME300000016