Weight Calibration - Excell PW Service Manual

Ip67 waterproof
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD

3-4 Weight Calibration

In the weighing mode, press the
Advanced Function setting mode. The LCD shows
or
key to select
UNITS
ACTIONS
Press
key
TARE
When the digit on the right
flashes, ensure that the
weigh pan is empty. Then
Press the
TARE
key.
When stable
Place the displayed weight
on the weigh pan. Press the
key. The scale
TARE
calculates the full capacity
internal value and after
stabilization, the buzzer
beeps 3 times. Unload the
weight pan and the
procedure
is completed.
In approved models,
n
Weight calibration conditions:
n
The calibration weight value placed on the weight pan must be over 100d, and the standard
deviation of the weight must be within 10%.
PW Service Manual
and
NET/GROSS
to enter the weight calibration mode.
DISPLAY
kg
.
kg
kg
.
set as 1 or 3, then
42
keys at the same time to enter the
ON/ZERO
NOTE
The LCD displays the calibration
weight value (the maximum capacity
st
of 1
unit) and the unit. The digit on
the right side keeps flashing.
The scale is calculating the internal
value of the zero point. After
stabilization, the LCD shows the
calibration weight value.
ON/ZERO key ⇒ Increment number by 1
UNITS key ⇒ Decrement number by 1
TARE key ⇒ Move cursor one place to the right
NET/GROSS key ⇒ Move cursor one place to the left
is disabled.
and use the
NET/GROSS
SME300000016