2-2-1 Weight Calibration Setting

For the most accurate weight calibration it is recommended that the indicator is calibrated at
n
full load where possible, although calibration at less than full load is also possible.
Press the
key
TARE
Ensure the weigh pan is empty
Press the
key
TARE
When stable
Place full load (30kg in this
example) on the weigh pan
then press the
key
TARE
Remove the weights from the
weigh pan and press any key
to exit.
Input "00"
Press the
key to
TARE
complete weight calibration
PW Service Manual
The display shows the calibration weight
kg
value and the weight unit set at the first
.
position
4 The calibration weight value can be
changed if required
kg
The indicator calibrates the zero point.
kg
When the weight is stable, the display
shows the calibration weight value.
.
The indicator shows the weight value.
kg
The buzzer sounds "beep, beep, beep"
.
when stable. Unload the weigh pan.
ON/ZERO key ⇒ Increment number by 1
UNITS key ⇒ Decrement number by 1
TARE key ⇒ Move cursor one place to the right
NET/GROSS key ⇒ Move cursor one place to the left