Omega DP41-S User Manual page 94

High performance strain gage indicator
Hide thumbs Also See for DP41-S:
Table of Contents

Advertisement

8
Scaling to
Display
Engineering
Units
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF
L2CNF
L3CNF
L4CNF
INPUT
BRIDGE
RDG.CNF RDG.1=0
RDG.2=0
RDG.3=0
RDG.4=1
RDG.5=
RDG.6=
RDG.7=0
RDG SC
1.00000
RDG OF
0000000
IN CNF
INP.1=0
INP.2=0
INP.3=0
INP.4=0
INP.5=
INP.6=0
INP.7=1
IN.SC.OF
DEC PT
FFFFF.F
CNT BY
Skip past
Skip past
Skip past
Skip past
Meter's input
2-point scaling IN.SC.OF
Active decimal point
Normal display brightness
Leading zeros suppressed
Disregard not used
Disregard not used
External hard reset vs
only peak reset
Preset for IN.SC.OF
Preset for IN.SC.OF
60 Hz ac power
Slow reading
(S2A jumper omitted)
Unipolar input
(S2B jumper omitted)
Std, for current inputs
Disregard not used
Disables IN.SC.OF
Ratiometric input
Skip past
By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point
will be entered later)
Press the 'RESET' two
times to return to the
run mode
8-50