9.2 METER FUNCTION MENUS

9.2.1 Input

By pressing the 'MAX' and 'MENU' buttons you can select "BRIDGE".
Refer to the chart below to go the next level of programming of the meter.
NOTE: Current, voltage or potentiometer inputs and their respective signal
input ranges are selected via the push-on jumpers located at S2 on the top
of the meter. Refer to Section 6 for a detailed explanation of these setting
and signal ranges.

9.2.2 RDG.CNF (Reading Configuration)

Reading configuration is used to select:
• reading scale and offset (direct vs 2-point) [RDG.1]
• active or independent decimal point [RDG.2]
• display brightness [RDG.3]
• leading zero suppression on your meter display [RDG.4]
Direct scale and Offset: these two values are used in the straight line
equation, y = mx + b. (RD.SC.OF)
Display = m times input plus b or [ m (input) + b ] (where m is the RDG SC
and b is the RDG OF).
RDG.6 enables or disables (RD.SC.OF), (RDG,SC), (RDG.OF).
The 2-data-point method allows the user to use two known points to convert
from one scale to another. For example, to convert from degrees Fahrenheit
to degrees Celsius, enter two (2) known points, such as 32°F = 0°C and
0°F = 17.77°C. The meter will automatically compute scale and offset and
display the correct value.