9.2.13 AL RDG (Alarm Readings)

Refer to Section 10 for an in-depth discussion of these features.
Alarm reading is used to select the number of readings (from 01 to 15)
the meter must make prior to activating the alarms.
MENU
BUTTON
MAIN MENU
"AL RDG":
* Press the 'MIN' button to show the current number of readings on the
display (left pair of digits are flashing).
The 'MAX' button is used to change the value of the flashing digits (from 01
to 15).
Pressing the 'MIN' button allows you go to the second set of digits.
The 'MAX' button is used to change the value of the flashing digits.
After changing the last number, if necessary, pressing the 'MENU' button
stores everything that was changed and advances the meter to the next
configuration ("SP DB").
The underlined item is the factory preset value.
9-22
MIN/MAX/MENU *
BUTTON
SUB MENU
"00 00"
("03 03")
DESCRIPTION
ALARM NUMBER
OF READINGS
Number of readings to
delay activation of Alarms
1 & 2 (from "01" to "15")
(AL1), (AL2)