9.2.17 OT.SC.OF (Output Scale and Offset)

Output scale and offset is used to calibrate your optional analog output to
correspond to the engineering units you desire.
MENU
BUTTON
MAIN
MENU
"OT.SC.OF":
* The 'MIN' button allows you to sequence through "READ 1", "OUTPT1",
"READ 2", and "OUTPT2" headings.
The 'MAX' button sends you to the value corresponding to "READ 1",
"OUTPT1", "READ 2", and "OUTPT2" so you can change it (go to the SUB
MENU 2 item).
** The 'MIN' button allows you to step through the digits of the applicable
number being changed.
The 'MAX' button changes the value of the digit to be displayed.
The 'MENU' button stores the selected values for each input required in
"OT.SC.OF". After the last value ("READ 2") has been entered and the
'MENU' button is pressed, the meter display will advance to "BAUD".
Every underlined item is the factory preset value.
Meter Function Menus
MIN/MAX/
MIN/MAX *
MENU **
BUTTON
BUTTON
SUB MENU 1 SUB MENU 2 DESCRIPTION
READ 1
000000.
("000000.")
OUTPT1
00.0000
("04.0000")
READ 2
000000.
("100000.")
OUTPT2
00.0000
("20.0000")
OUTPUT SCALE & OFFSET
Item #1 of Coordinate #1.
Enter the first value displayed
by the meter.
Item #2 of Coordinate #1.
Enter first desired
output value.
Item #1 of Coordinate #2.
Enter the second value
displayed by the meter.
Item #2 of Coordinate #2.
Enter second
desired output value.
9
9-25