Omega DP41-S User Manual page 83

High performance strain gage indicator
Hide thumbs Also See for DP41-S:
Table of Contents

Advertisement

READING SCALE (RDG SC) vs DECIMAL POINT SHIFT
NUMBER OF SPACES
DECIMAL POINT MOVES
TO THE LEFT
0
1
2
3
4
Go back into the setup mode and enter the new "RDG SC" value,
the 2-point calibration data and move the decimal point.
You must proceed as follows:
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF
L2CNF
L3CNF
L4CNF
INPUT
RDG.CNF
RDG SC
0.00100
RDG OF
IN CNF
INP.6=1
IN.SC.OF INPUT 1
READ 1
INPUT2
READ 2
DEC PT
FF.FFFF
CNT BY
TABLE 8-4
RDG SC
VALUE
1.00000
0.10000
0.01000
0.00100
0.00010
Skip past
Skip past
Skip past
Skip past
Skip past
Skip past
Enter new scale value
from above.
Skip past (Leave as 000000)
Activates "IN.SC.OF"
(2-point scaling)
20054.3
This is where you
00000.0
enter the data
66675.6
obtained during the
02000.0
calibration
run (see above)
Correct decimal point
selection
Press the 'RESET' button
two times to return to the
run mode
8-39
8
Scaling to
Display
Engineering
Units