6.4 JUMPER SETTING(S) FOR SENSOR EXCITATION

1). Maximum excitation for "Ratiometric" measurement is 10Vdc.
Note
2). Remove S2T jumper when you use external sensor excitation.
6.4.1 JUMPER SETTING(S) FOR SENSOR EXCITATION- ac Powered Unit
Your ac-powered meter is capable of supplying either 1.5 to 11Vdc or 24 Vdc
sensor excitation. (refer to Figure 6-2.).
6-3
24 Vdc meter excitation (S2N, S2T, & S4B)
S2
A
A B
S1
C
10 Vdc meter excitation (S2N, S2T, & S4A, S4C)
S2
A
A B
S1
C
1.25 Vdc meter excitation (S2Q)
S2
A
A B
S1
C
160 µA meter excitation (S2P)
S2
A
A B
S1
C
1.6 mA meter excitation (S2P, S2V)
S2
A
A B
S1
C
For 1.5 to 11Vdc excitation, install S4A and C jumpers, then adjust the
potentiometer (R39) for proper voltage.
For 24Vdc excitation, remove S4A and C jumpers and install S4B.
B
E
K
M
D
G
H
P
L
J
N
Q
F
B
E
K
M
D
G
H
P
L
J
N
Q
F
B
E
K
M
P
D
G
H
L
F
J
N
Q
B
E
K
M
H
P
D
G
L
F
J
N
Q
B
E
K
M
H
P
D
G
L
F
J
N
Q
R
S
T
U
V
R
S
T
U
V
R
S
T
U
V
R
S
T
U
V
R
S
T
U
V

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dp41-s-230Dp41-s-gnDp41-s-230-gnDp41-s-dc-gnDp41-s-dc

Table of Contents