Figure 8-10 S3 And S4 Main Board Jumper Positions - Omega DP41-S User Manual

High performance strain gage indicator
Hide thumbs Also See for DP41-S:
Table of Contents

Advertisement

A. Set the correct jumpers on the meter as shown in Figure 8-10.
(Refer to Section 6 for additional information)
Configure jumpers on S1 and S2 as shown in Section 6.
Install jumpers S3A & S3C and S4A & S4C on the main board.
S4A & S4C will provide a factory set 10 Vdc. Adjusting R38 will
vary excitation. Use a multimeter when adjusting R38.
DISPLAY
BD
R38
S4
A
B
C
W2
AC POWER BD
Figure 8-10. S3 and S4 Main Board Jumper Positions
Install jumpers on S1 and S2 as shown in Figure 8-11.
(FACTORY CONFIGURED)
C
B
A
S3
T1
W3
W1
8-41
SIGNAL INPUT BD
MAIN BD
J10
J2
J1
8
Scaling to
Display
J9
Engineering
Units