Advertisement

Error code or
Source
problem
E61
Temp
E63
Temp
E66
Temp
E--
Temp
E80
NIBP
E82
NIBP
E83
NIBP
E84
NIBP
E85
NIBP
E89
NIBP
*Acknowledging an alarm by pressing the Silence button, cancels and resets the alarm condition.
5-6
Troubleshooting: Alarm code interpretation
Definition
Temp probe broken (Alaris only)
Temp probe disconnected (Alaris only)
Temp probe too hot (Alaris only)
Temp scanner error (Exergen only)
Overpressure
Excess air in cuff
NIBP pump timeout
NIBP total timeout
NIBP level timeout
NIBP no determination
CARESCAPE V100 Vital Signs Monitor
Can be
acknowledged
Probable source
(silenced)?*
No
Replace temperature probe
Yes
Check for correct probe
Yes
Replace probe
Yes
Check Exergen battery, and
check for valid temperature
value. Also check the
scanner's display for
Exergen error information.
Refer to the
specific error codes"
for more information.
Yes
Excess cuff pressure. Check
for pinched or occluded
internal tubing.
Yes
Determination cannot be
made due to an excess
amount of air in the cuff.
Yes
Leak in cuff or o-ring in
hose
Internal leak in tubing or
pneumatic valve
Pump not turning on
Check or replace hose or
cuff.
Yes
Length of time has
exceeded 2 minutes for an
adult/pediatric
determination or 85
seconds for a neonate
determination.
Yes
Remained at one cuff
pressure level for more than
1 minute.
Yes
Cuff placement on patient
"Exergen-
section
2048723-002C

Hide quick links: