GE CareScape V100 Service Manual page 88

Vital sings monitor
Hide thumbs Also See for CareScape V100:
Table of Contents

Advertisement

Heart rate reading (bpm)
NIBP overpressure
'E80' displayed on Systolic display
Pump will not start
Overpressure alarm can be cleared
Display
All 7-Segment LEDs Light Correct Color
All Discrete LEDs Light, Correct Color
External DC detection
CHARGING indicator LED illuminated
Temperature test
Alaris
Temperature reading at 80.2° F
Temperature reading at 98.6° F
Temperature reading at 107.8° F
Exergen
Temperature reading of reference instrument
Temperature reading of instrument being verified
Difference between temperature reading of reference
instrument and instrument being verified
SpO
2
Pulse Value Displayed
Saturation Value Displayed
Signal Strength Bar
'---' displayed on SpO
Alarm is silenced, error display remains
Pulse Value
Saturation Value Displayed
Signal Strength Bar
Printer test
Printout is generated cleanly
Safety testing
Patient leakage
Fwd Pol, Neut Clsd, Gnd Clsd
Fwd Pol, Neut Open, Gnd Clsd
Fwd Pol, Neut Clsd, Gnd Open
Rev Pol, Neut Clsd, Gnd Open
4-30
Maintenance: Test results form
Description
display
2
CARESCAPE V100 Vital Signs Monitor
Min
Max
79.9° F
80.5° F
98.4° F
98.8° F
107.5° F
108.1° F
±0.2° C
(0.4° F)
100µA
500µA
500µA
500µA
Actual
Pass-Fail-N/A