GE CareScape V100 Service Manual page 95

Vital sings monitor
Hide thumbs Also See for CareScape V100:
Table of Contents

Advertisement

Error code or
Source
problem
976
System
979
System
980
System
984
System
985
System
986
System
989
System
994
System
999
System
E10
Printer
E11
Printer
E13
Battery
BATTERY
Battery
LOW
E00
Battery
---
SpO
2
E20
SpO
2
E21
SpO
2
E25
SpO
2
2048723-002C
Troubleshooting: Alarm code interpretation
Definition
Power supply voltage has peaked
above 18 Volts (incorrect power
supply)
Unknown power processor mode
received on power-up
Heap memory exhausted
Unused vector called
RTK 400hz timer re-entry
RTK 50hz timer re-entry
RTK overrun
Stack overflow
Background task stalled
Printer no paper
Printer too hot
Main battery low
Main battery is running low
Memory lost
SpO
Sensor off finger
2
SpO
sensor disconnected
2
SpO
replace sensor
2
SpO
lost pulse
2
CARESCAPE V100 Vital Signs Monitor
Can be
acknowledged
Probable source
(silenced)?*
No
External power brick issue
No
Mains PWA issue
No
Mains PWA issue
No
Return unit for evaluation
No
Return unit for evaluation
No
Mains PWA issue
No
Mains PWA issue
No
Mains PWA issue
No
Mains PWA issue
Yes
No paper in printer
Printer problem
Yes
Printer problem
Yes
Battery too low to
operate the unit
Charge battery
External DC source failed
Replace battery
Yes
Plug monitor in to recharge
Yes
Usually noted after
changing batteries.
User settings and date/
time revert to default
settings.
All entries in clinical
history are erased.
For units with Ohmeda
TruSignal technology,
verify the Line Frequency
(LF) mode setting. Refer to
"SpO2 configuration
settings"
Yes
Reposition SpO
Yes
Sensor disconnected
Yes
Replace SpO
Yes
Reposition SpO
on page 3-9.
sensor
2
sensor
2
sensor
2
5-5