2048723-002C
Installation: Configuring the monitor
NOTE
The first tone sounds as you first press the Menu button and the second
tone sounds after the monitor has saved the settings.
After all options pages have been displayed, the display will return to the first
service mode page (initial calibration page). The Service Mode option pages are
as follows:
Displayed on monitor
0 (in min window)
1 (in min window)
2 (in min window)
3 (in min window)
3 (in min window)
4 (in min window)
CARESCAPE V100 Vital Signs Monitor
Function
Refer to calibration section for functions.
Refer to calibration section for functions.
Refer to calibration section for functions.
NIBP Algorithm Type loaded
(Displayed in MAP/Cuff display window)
1: DINAMAP Classic NIBP
2: DINAMAP Auscultatory NIBP
3: DINAMAP SuperSTAT NIBP
NOTE
Changing setting effects NIBP performance.
Alarm mode
(Displayed in Pulse Rate window)
0: IEC mode
1: Legacy mode
NOTE
Use the Alarms button to toggle between
the available options.
Refer to the "Alarms" section of the
operator's manual for a detailed
description of the alarm modes.
SpO
Type loaded
2
(Displayed in SpO2% display window)
0: No SpO
2
1: Nellcor
2: Masimo
3: Ohmeda
NOTE
Incorrect setting will cause fatal 930 alarm
during operation.
3-13