Installation: Configuring the monitor
NOTE
For monitors configured for Ohmeda TruSignal SpO
setting for Line Frequency mode (LF) is correct for your country. Refer to
"SpO2 configuration settings"
SpO
configuration settings
2
Procedure for units with Ohmeda TruSignal technology
1.
With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2.
Press the Menu button until LF appears in the Pulse Rate window.
3.
Use the +/- buttons to select the option.
4.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.
WARNING
The Line Frequency (LF) mode must be set according to each
country's electrical power utilities implementation. The LF
mode must be checked and reset any time the monitor is set to
or reverts to factory default settings.
If the LF mode is set incorrectly, the susceptibility to ambient
light is increased and low perfusion performance may be
affected, resulting in potentially inaccurate readings.
Procedure for units with Nellcor technology
1.
With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2.
Press the Menu button until n0d (response mode) appears in the Pulse Rate
window.
3.
Use the +/- buttons to select the option.
4.
Press the Menu button once. SAt (SatSeconds) appears in the Pulse Rate
window.
5.
Use the +/- buttons to select the option.
6.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.
Procedure for units with Masimo technology
1.
With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2.
Press the Menu button until n0d (averaging time) appears in the Pulse Rate
window.
3.
Use the +/- buttons to select the option.
4.
Press the Menu button once. SAt (FastSAT) appears in the Pulse Rate window.
5.
Use the +/- buttons to select the option.
CARESCAPE V100 Vital Signs Monitor
only, verify that the
2
on page 3-9.