Installation: Configuring the monitor
6.
Press the Menu button once. SEn (sensitivity mode) appears in the Pulse
Rate window.
7.
Use the +/-buttons to select the option.
8.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.
Temperature unit of measurement configuration settings
Procedure for units with Alaris probes
1.
With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2.
Press the Menu button until Unt (unit of measurement) appears in the Pulse
Rate window.
3.
Use the +/- buttons to select the option.
4.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.
Procedure for units with Exergen scanner
The Exergen scanner comes preset with the requested unit of temperature
measurement, but can be changed. To change the scanner's unit of
measurement (°C or °F):
1.
Unplug the scanner from the monitor's host communication port.
2.
On the back of the scanner, loosen the single screw at the bottom and
remove the cover.
Cover
Single screw
CARESCAPE V100 Vital Signs Monitor
2048723-002C