LNOP
LNOP ADT
LNOP NEO
LNOP NEO-L
Foot
Finger
LNOP NEO PT-L
LNOP Adtx
LNOP Pdtx
LNOP DCI
LNOP DCIP
LNOP Hi Fi-Neo/adult
Foot
Finger
LNOP Hi Fi-Infant/Ped
LNOP Blue Infant Thumb/Toe*
LNOP YI Multi-Site
Foot/hand
Finger/toe
LNOP DC-195
LNOP TC-I
LNCS TCI
2048723-002C
Technical specifications and default settings: Specifications
Specifications: Masimo sensor accuracy
Sensor model
CARESCAPE V100 Vital Signs Monitor
SpO
range
2
70% to 100%
± 2 digits without motion
± 3 digits without motion
± 3 digits without motion
± 2 digits without motion
± 3 digits without motion
± 2 digits without motion
± 2 digits without motion
± 2 digits without motion
± 2 digits without motion
± 3 digits without motion
± 2 digits without motion
± 2 digits
± 3 digits (for 80-100) without motion
± 4 digits (for 60-80) without motion
± 3.3 digits (for 70-100) without motion
± 3 digits without motion
± 2 digits without motion
± 2 digits without motion
± 3.5 digits without motion
± 3.5 digits without motion