Sensor model
OxyTip++
TS-F-D
TS-E-D
TS-W-D
TS-SE-3
TS-AF-10
TS-AF-25
TS-F4-GE
TS-F2-GE
TS-E4-GE
TS-E2-GE
SpO
Sensor model
70% to 100%
TS-SA4-GE
±2 digits
TS-SA-D
Sensor light source
Wavelength*
Average power
* Information about wavelength range can be especially useful to clinicians.
A-6
Technical specifications and default settings: Specifications
Specifications: Ohmeda sensor accuracy
Specifications: Ohmeda sensor accuracy
range
SpO
range
2
2
90% to 100%
±1 digits
Specifications: Ohmeda sensor accuracy
CARESCAPE V100 Vital Signs Monitor
SpO
70% to 100%
±2 digits without motion
±3 digits without motion
±2 digits without motion
±2 digits without motion
±2 digits without motion
±2 digits without motion
±2 digits without motion
±2 digits without motion
±3 digits without motion
±3 digits without motion
SpO
range
SpO
2
80% to 90%
70% to 80%
±2 digits
±3 digits
Infrared: 930 to 950 nm (nominal)
Red 650 to 670 nm (nominal)
< 1 mW
range
2
range
SpO
range
2
2
below 70%
unspecified
2048723-002C