4-28
Maintenance: Safety testing
4.
Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – CLOSED
5.
Read the current leakage indicated on the tester and record the results on
the
"Test results form"
6.
Remove the AC mains voltage from the device under test.
7.
Repeat the steps in this procedure using the appropriate Test Body for any
additional patient connectors requiring safety testing.
If measured reading is greater than the acceptance criteria below, the device
under test fails. Contact GE Technical Support.
Acceptance criteria
All readings must be less than or equal to 5 mA.
CARESCAPE V100 Vital Signs Monitor
on page 4-29.
2048723-002C