3-8
Installation: Configuring the monitor
and select the parameter function. To change alarm settings, press the
Alarms button.
Inflation pressure default setting
Procedures
1.
Enter the configuration mode: with the monitor off, press and hold the Menu
button at the same time as pressing and holding the On/Off button for 3
seconds.
2.
Use the +/- buttons to increment or decrement the inflate pressure default
setting.
3.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press Menu.
Vital sign alarm default settings
Procedures
1.
Enter the configuration mode: with the monitor off, press and hold the Menu
button at the same time as pressing and holding the On/Off button for 3
seconds. After the unit enters the configuration mode, press Alarms. At any
point in the configuration mode menu, Alarms default can be selected.
2.
To set or change the default setting, press the Alarms button to select alarm
setting. Use the +/- buttons to increment or decrement the individual
settings.
NOTE
For the Alarms default setting to be saved, you must advance the menu
through the SpO
2
3.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press Menu.
Reverting to the factory default vital sign alarm limits
WARNING
The Line Frequency mode (for Datex-Ohmeda oximetry) must
be set according to each country's electrical power utilities
implementation; and it must be checked and reset any time the
monitor is set to or reverts to factory default settings.
To revert to the factory default vital sign alarm limit settings, the monitor must
be disconnected from the DC power supply and from the monitor battery. Refer
to
"Replacing the main battery on the monitor"
and battery disconnection/reconnection instructions.
When reverting to factory default settings, the user settings (including alarm
limits and inflation pressure), date/time, and the Ohmeda TruSignal SpO
Frequency mode (LF) will go back to default values. Refer to
settings"
on page 3-5 to configure the factory default user settings.
CARESCAPE V100 Vital Signs Monitor
settings.
on page 4-8 for DC power supply
Line
2
"Configuration mode
2048723-002C