Parts lists and drawings: Field-replaceable units (FRUs)
1
2
3
4
5
6
7
8
9
FRU 2037103-022: CARESCAPE V100 all hardware kit
NOTE
Each Hardware kit includes the following:
Item
1
Cable Tie, Locking, Nylon 6.6, 3.1" X 0.09"
2
Tie Wrap Intermediate, 2"
3
Screw, Mach Pnhd 832 X 2-1/2 SS
Vibratite
4
Screw, No6, Torx Pan, 1.0 Inch Vib
5
Screw, No8, Pozi-Pan 0.75 in Vib
6
Screw, M3 X 12, Self-Tap, Torx
7
Screw, Self Tap,Torx, Zinc
8
Kit, Screwlock Female With Thread-Lock
9
Screw, Mach Pnhd Phil 440 X 3/16 SS
Vibratite
10
Standoff Masimo SpO
11
Standoff Nellcor SpO
12
Standoff Ohmeda SpO
13
Wshr, #10 Fender Washer 1.00" OD
Stainle
14
Spacer, Steel Zinc
15
Foot, Rnd, 12.7 Dia X 3.5 H Self Adhsv
CARESCAPE V100 Vital Signs Monitor
20
13
14
15
16
17
10
11
12
18
Description
2
2
2
19
21
22
Torque
Qty
specifications
5
10
8 in-lb ± 0.5 in-lb
6
8 in-lb ± 0.5 in-lb
4
8 in-lb ± 0.5 in-lb
20
6 in-lb ± 0.5 in-lb
20
6 in-lb ± 0.5 in-lb
23
6 in-lb ± 0.5 in-lb
3
3 in-lb ± 0.5 in-lb
20
4
4
4
4
20
20
2048723-002C