P- 2.0.1 SYSTEM INITIALIZATION DOES NOT REACH 100%
IMPORTANT!
a)
Turn the machine OFF!
IMPORTANT!
b)
Reconnect the ACTUATOR cable.
Figure below
c)
, behind the card cage, unplug the
Pressure module!
d) From here forward, ignore "Art BP No Comm" OR "Blood Still Sensed!" if they occur!
e) Turn the machine ON (fan running). Does System Initialization
reach 100%?
Yes
System Initialization reaches 100%! See ISOLATING THE MODULES this page
No
System Initialization does NOT reach 100%! ENSURING the ACTUATOR cable has been
reconnected, see procedure number
ISOLATING THE MODULES
Turn the machine OFF!
A)
B) Your choice, CAREFULLY* plug one of the modules back in. *NEVER reverse connect the modules!
NEVER plug the Blood Pressure module in with the machine on!
C) Turn the machine on (fan running).
D) Repeat parts A through D until System Initialization does NOT reach 100% indicating the last module
plugged in (or its cable) is causing the problem. If after plugging in all modules and System Initialization
reaches 100% the problem may be heat related and requires running the machine in Heat disinfect to
locate it.
ALL MODULES ONLY
P- 2.0.2
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
including the Blood
(page 640).
Page 639