UF- 3.0.0 INADEQUATE UF PUMP OUTPUT

a) Per the Figure right, trace the UF Pump's cable
to distribution board position "P22, UF-P".
b) ENSURE the connector is placed
properly AND the cable is not damaged.
IMPORTANT, if NOT already, place
c)
the machine into RINSE!
d) Ensure the external flow indicator's 'bob' is moving up and down!
e) Per the Figure right, remove the tubing
from the UF Pump white INPUT (bottom)
nozzle. Is there flow through Filter #74
(Yes or No)?
Yes
Flow through the filter!
See procedure number
UF- 4.0.0
No
No flow! This may be the
problem! Replace UF Filter
#74.
Figure 87 – UF Pump Springs / Seals/ Diaphragm
page 440
(page 507).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H