Advertisement

P- 2.0.0 INITIALIZATION DOES NOT COMPLETE / ISOLATE PUMPS
IMPORTANT! To prevent damage, turn the machine OFF!
a)
unplug the ACTUATOR CABLE
b) Figure below
c) Turn the machine on (fan running). Does System Initialization
reach 100%?
Yes
System Initialization reaches 100%! See ISOLATING THE PUMPS this page.
No
System Initialization does NOT reach 100%! See procedure number
ISOLATING THE PUMPS
Possible 24 volt 'short circuit' at the Acid, Bicarb, Heparin OR UF Pump.
IMPORTANT! Turn the machine OFF and reconnect the ACTUATOR cable.
A)
B) Figures below, from the distribution board, unplug a pump (your choice) then continue to part C to see if
this is the problem.
C) Turn the machine on (fan running). Does System Initialization reaches 100%?
Yes
System Initialization reaches 100%! The unplugged pump may be causing the problem! To
confirm this turn the machine off, plug the pump back in, and turn the machine on. If System
Initialization does NOT reach 100% the pump is bad.
No System Initialization does NOT reach 100%! Turn the machine off and repeat parts B and C until
System Initialization DOES reach 100%. If after ALL pumps have all been unplugged and System
Initialization still does NOT reach 100% either the Actuator Cable OR the distribution board may
be bad.
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
from the distribution board.
P- 2.0.1
(page 639)
Page 638