AIR- 1.0.3 AIR AT L1 / CHECK FOR AIR INPUT CHAMBER #69
Figure below, ensuring TMP is between 300 and 360 mmHg, for two (2) minutes, watch through the INPUT
rear side
tubing at the
of Chamber #69 (Location #10 (L10)). Air bubbles seen?
Yes
Air seen at L10! See procedure number
AIR-1.0.4
(page 494).
No air seen at L10! Return to the procedure that brought you here.
Figure 77 – Leak Checks (2) / Abbreviated Flow Diagram
2008K/K2 Troubleshooting Guide
Page 493
P/N 507298 Rev. H