Over- 6.0.0 Massive Overflow (More Than 600 Ml Per Minute) - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

OVER- 6.0.0 MASSIVE OVERFLOW (More than 600 ml per minute)

a) With the machine OFF, perform
b) Figure right, unplug the Deaeration Pump #20 from
distribution board position "P20, Degas-P".
c) If (and ONLY if) an "Actuator Board No Echo" OR "Failed
Sending Data To Actuator Board" banner EVER appear
there may be a problem with the Actuator-Test Board!
IMPORTANT! Turn the machine on!
d)
e) Per
Figure 103
(below), ENSURE the Deaeration Motor is
NOT rotating i.e. is unplugged!
Figure 103 – Hydraulics FRONT View / Deaeration Motor
IMPORTANT
f)
! With the machine AND water ON allow thirty (30) seconds. TWO (2) possible scenarios
1) or 2) below:
1)
IF (and ONLY if) massive overflow continues: With the machine AND water ON, see
procedure number
2)
IF massive overflow STOPS: Perform parts a AND b below:
a) AVOIDING the VACANT positions, return the
Deaeration Pump's connector to distribution board
position "P20, "Degas-P". The Deaeration Motor
shaft rotates if plugged in properly!
b) Per
Figure 103
motor shaft rotating COUNTERCLOCKWISE
(CCW) (Yes or No)?
INITIAL CHECKS
OVER- 7.0.0
(page 550).
(above), is the DEAERATION
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
(page 6) and return here if a problem is NOT located.