P- F.6.0 CHECK TRANSFORMER OUTPUT

a) Turn the machine off and UNPLUG it.
Electrocution hazard if not unplugged!
b) Slide the power supply away from the cabinet.
c) ENSURE all cables are plugged in securely! If not,
this may be the problem!
d) Remove the two front screws, un-mount the Power
Control Board from its four plastic clips
e) Lay the supply's rear panel down to access to the
rear side of the board.
f)
Plug the machine in.
IMPORTANT! Turn the machine ON (fan running)!
g)
Per the Figure below
h)
connector. More than 23.0 volts DC (Yes or No)?
Yes
More than 23.0 volts! See parts a AND b below:
a) Turn the machine off and UNPLUG it. CAUTION! Electrocution hazard if not
unplugged!
b) Bad thermal (BROWN) fuse S13 (16 Amp). To LOCATE S13 refer to
651).
No
Less than 23.0 volts! See procedure number
Figure 116 – Power Control Board
(ST12)
CAUTION!
CAUTION!
High voltage now present!
, measure from the rear (solder) side of the Power Control Board's ST12