Tmp- 2.0.5 Filact Was Or Is = 1; Tmp- 2.0.6 Isolate Uf Pump Control/Uf Pump Leak - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

TMP- 2.0.5 FILACT WAS OR IS = 1

This procedure ENSURES Valve #43 is not sticking closed AND allows Conductivity to increase:
IMPORTANT!
a)
From the Home screen, select the [Dialysate Flow] window!
b) Set [Dialysate Flow] to 800 ml/min and press 'CONFIRM'!
c) Call debug screen 0. WITHOUT LOOKING AWAY, watch Flow
Error for four (4) minutes. It should NEVER = 1! TWO (2)
possible scenarios:
1)
IF (and ONLY if) Flow Error EVER = 1 , even if only once: Proceed to
1, FLOW ERRORS IN DIALYSIS
2)
IF Flow Error ALWAYS = 0! See parts a THROUGH e below
a) From the Home screen, allow [Conductivity] to stabilize between 13.0 and 14.5 mS!
IMPORTANT! Open the shunt door and LEAVE IT OPEN till instructed!
b)
Install a 'dummy venous chamber' in the Level Detector module!
c)
d) Call debug screen 1. Allow FILACT = 0 ALWAYS. If FILACT DOES NOT = 0, after two (2)
minutes, most likely, the female Air Sensor connector is NOT plugged in PROPERLY!
e) ALARM RESET SEQUENCE: i) Press and release the 'Reset' key then immediately press
and hold it for three (3) seconds; ii) Allow thirty (30) seconds. If ANY alarm reoccurs attempt
RESET up to twice more BEFORE continuing to procedure number
TMP- 2.0.6 ISOLATE UF PUMP CONTROL / UF PUMP LEAK
a) ENSURE the lamp above the front panel's UF on/off key is
OFF!
b) Call debug screen 0.
Figure right
c)
,
WITHOUT LOOKING
AWAY, watch the UF
Pump's 'dot' for forty five
(45) seconds. It MUST
REMAIN white! Does it
EVER 'blink' to blue (Yes
or No)?
PROGRAM.
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 20
TMP- 2.0.6
OFF
i.e. UF is
,
SECTION
(page 452).
Page 452