TMP- 2.0.0 DEAERATION PRESSURE OKAY

Turn the water OFF!
a)
b) Figure right, ENSURE the external flow
indicator's inner tapered tube is NOT
scored (i.e. tube really 'cloudy'). If
scored, replace it!
c) Allow a "No Water" alarm to occur
THEN, after one (1) minute, the
deaeration motor stops.
d) Remove the Deaeration gauge and
reattach the tubing!
IMPORTANT! Turn the water on and allow the "No Water" alarm to go away!
e)
f)
Remove the hydraulic compartment from the cabinet!
g) Return BOTH concentrate connectors to their rinse ports.
Place the machine into RINSE to pressurize the secondary hydraulic circuit!
h)
From here forward, ENSURE a "No Water" alarm does not occur!
i)
Unless as external leak is immediate allow ninety (90) seconds minute BEFORE continuing to
procedure number
TMP- 2.0.1
(page 447).